CO2Sense.net

CO2Sense.net mass spectrometer 

Description :

By using the IMR technique, the measurement becomes flexibly and easy to adapt to new applications, selective and free of any interference.Unlike others, no fragmentation or overlapping spectra can damage the interpretation of the detected results.

IMR means using primary ions with lower energy level (10 eV .. 14 eV) to completely ionise the probe gas molecules. The signal noise ratio will be optimised by the integrated Octopol separator focusing the primary ions and filtering out any interferences.The quadrupol mass filter (0-500 amu) than separates the molecules for further detection at the fast pulse counter.

The temperature and pressure compensated sample gas inlet guaranties the correct measurement and avoids discrimination of the gas. Any contamination due to condensation or particulate matters is minimized by the unique sample inlet design.